What OURA and WHOOP Heart Rate Averages Hide

An independent 483-person James Cook University study shows the lag and individual error that disappear when wearable heart-rate data are averaged

Based on Journal of Human Science & Technology, Volume 1, Issue 2, Article 7.[1]

The published study reports averages across 483 people. Those averages matter, but they are not what a wearer sees. A wearer sees one number from one ring or wrist strap at one moment. If that number is late or wrong, an accurate group average offers little comfort.

I went back to the participant-level records to show what the pooled curves hide. The five subjects below are real study traces selected for the largest combined persistent error across OURA and WHOOP during exercise. They are deliberately the most divergent cases, so they should not be read as typical performance. They show the kind of individual behavior that positive and negative errors can erase when everyone is averaged together.

This is an independent James Cook University study

This study was conducted and authored by Kai Weng and Ethan Wei of James Cook University, Singapore. The published paper reports partial support from the Australian National Health and Medical Research Council, and no funder role in the study design, data collection, analysis, interpretation, manuscript preparation, or publication decision.[1] My role in this blog is to interpret their published data and results for the reader.

Not the first heart-rate study—but a different scale and test

Heart-rate validation studies of OURA and WHOOP existed before this paper. Most of the widely cited OURA evidence examined resting or nocturnal averages: 49 adults in OURA-linked research,[3] 35 adults in an independent overnight study,[4] 53 adults in a one-night multi-device laboratory study,[5] and 13 adults followed across many nights.[6] WHOOP had also been assessed during sleep[7] and in smaller exercise cohorts.[8]

Dynamic exercise evidence was much thinner. A 2026 peer-reviewed comparison enrolled 45 adults, but only 10 wore an OURA Ring.[9] A Japanese exercise report evaluated one person (Oura Ring 3 vs Polar H10).[10] Apple subsequently released a manufacturer-run September 2026 white paper enrolling 1,460 people and testing OURA Ring 5 and WHOOP 5 in multiple activities, with at least 50 participants required for each device-by-workout comparison.[11]

The JCU investigation included 483 adults, with OURA Rings, and simultaneously worn WHOOP sensors, a standardized 30-minute treadmill run, 11 paired time points, and 5,313 paired observations per wearable.[1] Based on the published literature identifiable through November 15, 2026, it appears to be the largest independent academic study reporting both OURA and WHOOP against the same reference during standardized sustained treadmill running.

How the treadmill comparison worked

After warmup, 483 participants completed a 30-minute treadmill run at an eight-minute-mile pace. The study recorded the manual reference heart rate and the OURA and WHOOP outputs at baseline and every three minutes through minute 30. Each participant wore a personally owned, properly fitted OURA Ring with an established baseline. WHOOP was positioned on the underside of the wrist over the radial artery during the run.

The Deep Health® plantar platform was not used during treadmill running because it requires continuous ground contact. Its blood-pressure and heart-rate comparisons in the paper apply to stationary testing.[1]

The Apple Watch did not provide paired transition data

They also attempted to include the Apple Watch Series 10. When heart rate was changing rapidly, the watch repeatedly showed the spinning measurement indicator instead of returning a usable value at the scheduled time point. Some displays remained locked or unavailable for several minutes and produced a value only after the transition had passed or the run had stopped.

That made a simultaneous comparison impossible. The Apple iWatch arm was abandoned and excluded from the quantitative analysis. This was not a finding that its numerical value was high or low. The problem was more basic: during the transition we intended to measure, it often did not return a value that could be paired with the reference.[1]

What the pooled chart shows

At baseline, the cohort averages were close: the reference was 85.13 bpm, OURA was 85.40 bpm, and WHOOP was 85.38 bpm. Three minutes after the treadmill started, the reference had risen by 60.38 bpm to 145.51 bpm. OURA averaged 134.04 bpm and WHOOP 135.23 bpm. The mean errors were therefore −11.47 bpm for OURA and −10.28 bpm for WHOOP at the moment of rapid acceleration.[1]

Pooled treadmill charts show OURA and WHOOP reading 10–11 bpm below the manual reference at minute 3, with wide limits of agreement late in exercise.
Figure 1. Pooled results. The left panel shows the cohort-level lag at minute 3. The right panel shows the late-exercise 95% limits of agreement: −17.98 to +19.60 bpm for OURA and −16.97 to +17.12 bpm for WHOOP.

After minute 6, the pooled lines appear to converge. That visual convergence is misleading if it is interpreted as individual accuracy. During minutes 12 through 30, the mean signed error was only +0.81 bpm for OURA and +0.08 bpm for WHOOP, yet the individual mean absolute errors remained 8.37 and 7.56 bpm. Positive and negative errors were cancelling each other.

Only 30.2% of OURA readings and 34.3% of WHOOP readings fell within ±5 bpm of the reference during sustained running. Errors greater than 10 bpm occurred in 32.6% of OURA readings and 24.4% of WHOOP readings. For OURA, 8.4% of readings differed by more than 15 bpm.[1]

How the five individual cases were selected

For this blog, I calculated each participant’s mean absolute error across the ten active exercise readings from minute 3 through minute 30 for both OURA and WHOOP. I then averaged the two device-specific MAEs and selected the five highest combined values. This is a descriptive ranking of the supplied participant records. It is intended to expose user-level behavior, not to estimate how frequently these worst cases occur.[2]

Heart-rate errors in the five selected participants
Participant OURA error at 3 min WHOOP error at 3 min OURA exercise MAE WHOOP exercise MAE Largest observed error
882784 -20 bpm -14 bpm 12.6 bpm 13.8 bpm OURA -22 bpm at 30 min
129979 -18 bpm -12 bpm 11.9 bpm 11.9 bpm OURA -20 bpm at 9 min
935556 -21 bpm -10 bpm 11.4 bpm 12.1 bpm WHOOP +24 bpm at 15 min
935007 -13 bpm -11 bpm 12.1 bpm 11.4 bpm OURA -20 bpm at 30 min
642110 -8 bpm -16 bpm 12.3 bpm 10.5 bpm OURA -22 bpm at 24 min

Negative error means the wearable read below the manual reference. Positive error means it read above the reference. MAE is the average absolute error across minutes 3 through 30.

Five traces the average cannot show

OURA and WHOOP read below the manual reference at minute 3 in all five selected treadmill traces. WHOOP later reads above it in one trace.
Figure 2. Manual reference, OURA, and WHOOP trajectories for the five participants with the largest combined persistent error. The dotted vertical line marks minute 3.

Participant 882784

The reference reached 154 bpm at minute 3. OURA showed 134 bpm, 20 bpm low, and WHOOP showed 140 bpm, 14 bpm low. The under-reading persisted. At minute 30, OURA was 22 bpm below the reference and WHOOP was 21 bpm below it.

Participant 129979

At minute 3, the reference was 158 bpm. OURA read 140 bpm and WHOOP 146 bpm. OURA’s largest error was −20 bpm at minute 9, while WHOOP’s largest error was −17 bpm at minute 12.

Participant 935556

This trace shows why a single average can be deceptive. At minute 3, OURA was 21 bpm below the reference and WHOOP was 10 bpm below it. By minute 15, WHOOP had crossed to the other side and read 24 bpm above the reference. The direction of the error changed while the group average remained reassuring.

Participant 935007

Both devices under-read through most of the run. At minute 3, OURA was 13 bpm low and WHOOP 11 bpm low. At minute 30, OURA remained 20 bpm below the reference. WHOOP reached its largest deviation of −16 bpm at minute 27.

Participant 642110

The reference was 159 bpm at minute 3. OURA showed 151 bpm, while WHOOP showed 143 bpm. Later, OURA was 22 bpm below the reference at minute 24 and WHOOP was 17 bpm below it at minute 12.

OURA and WHOOP errors for five selected participants often fall outside the shaded ±5 bpm band. One WHOOP reading is 24 bpm above the manual reference.
Figure 3. Signed error for the same five participants. The gray band marks ±5 bpm. Values below zero are under-readings; values above zero are over-readings.

Why the display can look smooth and still be late

OURA and WHOOP estimate heart rate from optical changes in peripheral blood volume. Running adds arm movement, shifting skin contact, cadence-related artifacts, and changes in peripheral blood flow. The software must decide which pulses are real and which are noise. Acquisition windows, rejection rules, and smoothing can stabilize the displayed number, but stability can come at the cost of delay.

That tradeoff is most visible when heart rate is changing rapidly. The number on the screen may describe where the heart rate was rather than where it is. Later in the run, a cohort curve can appear to catch up even though individual devices continue to alternate between under-reading and over-reading.

Why this matters for women’s cycle features

This study did not test menstrual-cycle phase, ovulation, fertility, or contraception. It cannot be cited as direct proof that a cycle-prediction or birth-control feature fails. It does, however, expose an upstream measurement problem that matters whenever cardiovascular signals are used as inputs to a downstream prediction.

WHOOP states that Menstrual Cycle Insights combines logged periods with resting heart rate, heart-rate variability, skin temperature, respiratory rate, and other physiological data to estimate phases and adjust coaching.[12] OURA describes its own cycle and fertile-window features as using temperature trends together with physiological information that can include cardiovascular and respiratory signals.[13] If a model uses heart-rate-derived information, the accuracy, timing, and error handling of that input matter. A reassuring population average cannot establish that the input is correct for an individual woman.

Women were not a token subgroup in the JCU trial: 267 of the 483 participants were women. Exercise error was not materially smaller in women than in men. OURA exercise MAE was 8.56 bpm in women and 8.26 bpm in men; WHOOP MAE was 7.66 bpm in both groups.[1] Those findings justify demanding transparent, female-specific validation of any cycle feature that depends on cardiovascular measurements. They do not justify assuming that treadmill error is identical to nocturnal resting-heart-rate or HRV error; those data streams require their own validation.

There is also an essential birth-control distinction. OURA’s and WHOOP’s internal cycle features explicitly state that they are not contraceptive tools.[12][13] Natural Cycles is the FDA-cleared—not FDA-approved—Class II contraceptive software. Its formally declared inputs are temperature, menstrual dates, and optional ovulation or pregnancy tests.[14] OURA heart rate is not listed as a fertility-algorithm input in the current FDA summary, although Natural Cycles support documentation says that OURA can transfer additional sleep data, including heart rate, sleep length, and sleep stages.[15] That creates a legitimate need to disclose whether those additional signals are used by the regulated algorithm, for data-quality screening, or only as integration metadata.

Natural Cycles did not simply assume that OURA temperature was interchangeable with oral basal temperature. Its 2021 OURA integration clearance was supported by a study of 40 women across 223 cycles, including 155 complete cycles and 87 cycles with a positive luteinizing-hormone test. The algorithm detected ovulation using both OURA-derived and oral-thermometer inputs, and the OURA input produced 1.6 additional luteal-phase green days without an observed increase in unintended-pregnancy risk.[16] The current FDA framework also requires each newly integrated wearable to pass validation criteria focused specifically on false green-day risk.[14]

But clearance of the downstream algorithm does not make every upstream sensor value physiologically true. OURA measures distal finger-skin temperature and generates an overnight temperature trend; Natural Cycles then processes that trend into a single absolute temperature.[15] Safeguards can reject missing or implausible data and respond conservatively with more red days. They cannot reconstruct a true temperature from an undetected systematic sensor or transformation error. The 40-woman validation supports the combined OURA-to-Natural-Cycles pipeline under the tested conditions;[16] it does not establish that every nightly temperature is a direct basal-body-temperature measurement, nor does it validate OURA heart-rate accuracy. The JCU heart-rate findings therefore do not invalidate Natural Cycles, but they make the upstream measurement question impossible to dismiss.

What the user should take from this study

A pooled trend is not an individual measurement. The pooled curves show a clear early lag, but the individual records show something more consequential: the error does not simply disappear after the mean lines converge. For some users it remains large, changes direction, or grows later in the run.

Longitudinal data are not a substitute for measurement accuracy. Repeating a biased measurement merely creates a longer record of the bias. A hand-worn optical sensor operates within a position-dependent hydrostatic system. When the hand is level with the heart, the hydrostatic pressure difference is approximately zero. For every centimetre of vertical displacement, local vascular pressure changes by approximately 0.7 mmHg. A hand positioned 20 cm above the heart experiences an approximately 14 mmHg reduction, while the same hand positioned 20 cm below experiences an approximately 14 mmHg increase—a 28 mmHg difference across those two ordinary positions.

Arm movement can additionally alter tissue compression, venous pooling, peripheral perfusion, sensor contact and motion artefact. Consequently, an apparently smooth longitudinal trend may reflect limb position and algorithmic processing as much as the underlying physiology. A current heart-rate value must therefore be validated under the same posture, hand position, movement and transition conditions in which it will actually be used.

These hydrostatic effects matter because heart rate is inferred from a pressure-dependent peripheral pulse waveform rather than measured directly at the heart. Changes in local vascular pressure can alter pulse amplitude, signal quality, detection timing and the heart-rate value ultimately reported by the device.

The central lesson is straightforward: a smooth line is not proof of a current measurement, and an accurate group average is not proof that the number shown to one person is accurate.

Study and data

  1. Hemodynamic Fidelity Across Anatomical Acquisition Sites
  2. DOI and accompanying datasets (10.5281/zenodo.24176147)

The individual ranking and figures in this blog are a descriptive analysis of the participant-level workbooks accompanying the publication. Subject codes are de-identified study identifiers.